Objective: To compare the measurement of haemoglobin concentration ([Hb]) using the HemoCue haemoglobinometer with that using the Coulter STKS haemoglobinometer. Design: Thirty two EDTA samples were taken from neonates. [Hb] was measured in these samples using the HemoCue; the samples were then transferred to the haematology laboratory for [Hb] determination with the Coulter STKS. In addition, [Hb] was determined in 50 different random EDTA neonatal samples already held in the laboratory, using the HemoCue and Coulter STKS. Patients: Neonates in the intensive care and low dependency Units of the Royal Devon and Exeter Hospital. Interventions: Samples were collected from arterial lines or by venepuncture or heel prick into an EDTA bottle. Main outcome measures: [Hb] using the HemoCue and Coulter STKS methods. Results: The mean [Hb] measured using the HemoCue was 150.3 g/l (range 78-215) compared with 152.8 g/l (range 78-217) measured using the Coulter STKS, with a mean of the differences of 2.5 g/l. The standard deviation of the differences of the 82 samples was 3.73 g/l. The limits of agreement of the two methods (mean difference ± 2SD) were -4.8 to +9.8 g/l. Conclusion: With adequate training and monitoring, the HemoCue can be used directly on the neonatal unit for rapid determination of [Hb] to within 7.5 g/l compared with the laboratory Coulter STKS, using much smaller sample volumes.
CITATION STYLE
Rechner, I. J., Twigg, A., Davies, A. F., & Imong, S. (2002). Evaluation of the HemoCue compared with the Coulter STKS for measurement of neonatal haemoglobin. Archives of Disease in Childhood: Fetal and Neonatal Edition, 86(3). https://doi.org/10.1136/fn.86.3.f188