Delayed presentation of rupture after venous angioplasty

0Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Venous rupture is an uncommon complication resulting from dialysis graft interventions. We present a case in which contrast extravasation recurred following an initially successful balloon tamponade. The rupture site could not be negotiated with a guidewire to provide endovascular treatment, necessitating balloon occlusion of the graft. Copyright © 2007 by Thieme Medical Publishers, Inc.

Cite

CITATION STYLE

APA

Backer, M., & Zangan, S. (2007). Delayed presentation of rupture after venous angioplasty. Seminars in Interventional Radiology, 24(3), 324–326. https://doi.org/10.1055/s-2007-985744