Abstract
Background and Aims Positive pressure ventilation is a common intervention in neonatal resuscitation. Distraction, type of device and experience may influence performance. Studies have not included self-inflating bags (SIB) equipped with a PIP manometer and expiratory PEEP valve. We aimed to compare clinicians' ability to ventilate a mannequin using a SIB with additional manometers against a T-piece (TP), with and without distraction. Method 50 medical and nursing staff were tested using standarised case scenarios with a leak free intubated mannequin. Participants targeted PIP 30 cmH 2 O, PEEP 5 cmH 2 O, inflation rate (IR) 60 inflations/ minute with both devices in randomised order. We analysed PIP, PEEP, IR, expired tidal volume (TVe), professional group and compared devices during baseline and 3 minutes of distraction. Results 12,981 inflations were analysed. Mean (SD) ventilation parameters are shown in table. When analysed by operator, more variation was observed in IR (P=0.029) and TVe (P=0.002) with SIB during distraction. Conclusions Clinicians' general performance when using a SIB where PIP and PEEP are displayed is comparable to a T-Piece, however more variation in IR and TVe occurs under distraction. This may be relevant in a real resuscitation.
Cite
CITATION STYLE
Gupta, N., Thio, M., Dawson, J., Kamlin, C., Schmoelzer, G., & Davis, P. (2012). 401 Ventilation Parameters during Resuscitation: Comparison of two Different Devices in a Mannequin Model with and without Distraction. Archives of Disease in Childhood, 97(Suppl 2), A118–A118. https://doi.org/10.1136/archdischild-2012-302724.0401