Determination of uric acid in plasma by LC-MS/MS and its application to an efficacy evaluation of recombinant urate oxidase

22Citations
Citations of this article
33Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the determination of uric acid in human plasma was developed and validated. Separation was achieved on a C18 column by the mobile phase of 30% water (containing 0.5% formic acid) and 70% methanol. Quantification was done using a multiple reaction monitoring (MRM) mode to monitor the precursor-to-product ion transitions of m/z 169.1 → m/z 141.1 for uric acid and m/z 171 → m/z 143 for 1,3-15N uric acid (IS) at a positive ionization mode. The calibration curve was established over the range of 0.4096 - 100 mg/L, and the correlation coefficient was better than 0.99. The intra-day and inter-day relative standard deviations were less than 5.1%. The accuracy determined at three concentrations ranged between 92.7 and 102.3%. This method was successfully applied to an efficacy study of intravenous recombinant urate oxidase produced by Escherichia coli in a clinical phase I study. 2013 © The Japan Society for Analytical Chemistry.

Author supplied keywords

Cite

CITATION STYLE

APA

Luo, X., Cai, N., & Cheng, Z. (2013). Determination of uric acid in plasma by LC-MS/MS and its application to an efficacy evaluation of recombinant urate oxidase. Analytical Sciences, 29(7), 709–713. https://doi.org/10.2116/analsci.29.709

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free