Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing

27Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

This report describes the clinical management of 2 patients with ventricular fibrillation (VF) who received inappropriate shocks from an implantable cardioverter defibrillator (ICD) due to T-wave oversensing. Cardiac sarcoidosis was confirmed as the underlying heart disease in 1 patient and idiopathic dilated cardiomyopathy in the other. Within 2 months after ICD implantation, both patients received several inappropriate shocks during sinus rhythm. Stored electrograms showed decreased R-wave amplitudes and increased T-wave amplitudes. The ICD sensed both R- and T-waves as ventricular activation, which met the rate criteria for VF treatment. Reprogramming the sensing threshold in association with administration of a drug to slow the heart rate decreased the incidence of the inappropriate shocks in both patients, but these palliative measures did not completely suppress the inappropriate shocks. To avoid T-wave oversensing, the repositioning or adding of a sensing lead is required. The potential risk of T-wave oversensing in ICD patients who have small R-wave amplitudes should be recognized.

Cite

CITATION STYLE

APA

Washizuka, T., Chinushi, M., Kasai, H., Watanabe, H., Tagawa, M., Hosaka, Y., … Aizawa, Y. (2001). Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing. Japanese Circulation Journal, 65(7), 685–687. https://doi.org/10.1253/jcj.65.685

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free