Abstract
This report describes the clinical management of 2 patients with ventricular fibrillation (VF) who received inappropriate shocks from an implantable cardioverter defibrillator (ICD) due to T-wave oversensing. Cardiac sarcoidosis was confirmed as the underlying heart disease in 1 patient and idiopathic dilated cardiomyopathy in the other. Within 2 months after ICD implantation, both patients received several inappropriate shocks during sinus rhythm. Stored electrograms showed decreased R-wave amplitudes and increased T-wave amplitudes. The ICD sensed both R- and T-waves as ventricular activation, which met the rate criteria for VF treatment. Reprogramming the sensing threshold in association with administration of a drug to slow the heart rate decreased the incidence of the inappropriate shocks in both patients, but these palliative measures did not completely suppress the inappropriate shocks. To avoid T-wave oversensing, the repositioning or adding of a sensing lead is required. The potential risk of T-wave oversensing in ICD patients who have small R-wave amplitudes should be recognized.
Author supplied keywords
Cite
CITATION STYLE
Washizuka, T., Chinushi, M., Kasai, H., Watanabe, H., Tagawa, M., Hosaka, Y., … Aizawa, Y. (2001). Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing. Japanese Circulation Journal, 65(7), 685–687. https://doi.org/10.1253/jcj.65.685
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.