[Prospective electrocardiogram gated coronary 320-row area detector computed tomography angiography using low tube current scanning with full reconstruction].

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

It is possible to obtain equivalent image quality and a lower radiation dose using low tube current scanning with full reconstruction as compared to usual tube current scanning with half reconstruction in a 320-row area detector computed tomography (ADCT) angiography. Of 589 patient underwent coronary CT angiography (CCTA), 11 patients with (RR-PQ) ≥1069 ms were enrolled. In those patients, low tube current (50% mA) scanning with full or half reconstruction were performed. As a control, 11 patients with matched pairs of tube voltage, BMI and heart rate who underwent usual scanning with half reconstruction (100% mA with half reconstruction) were selected. Standard deviation of the CT value (SD) was measured in aorta (Ao), left atrium (LA) and left ventricle (LV), and extended dose-length products (DLP.e) were calculated. Significant motion artifact was not observed in any patients. SD of 50% mA with half reconstruction, 50% mA with full reconstruction, and 100% mA with half reconstruction were 28.1±2.6, 20.3±1.9, 20.7±2.5 HU in Ao, 34.4±4.4, 24.9±2.8, 24.9±3.1 HU in LA, and 29.7±2.3, 21.7±1.9, 22.1±2.3 HU in LV, respectively. There were not significant differences between 50% mA with full reconstruction and 100% mA with half reconstruction, but there were significant differences between 50% mA with half reconstruction and 50% mA with full reconstruction in all sites. The DLP.e of 50% mA scanning (74.1±21.8 mGy·cm) was significantly lower than 100% mA scanning (161.9±28.9 mGy·cm). CCTA with lower radiation dose and equivalent image quality can be obtained by ADCT using 50% mA scanning with full reconstruction in patients with (RR-PQ) ≥1069 ms.

Cite

CITATION STYLE

APA

Sano, T., Matsutani, H., Kondo, T., Fujimoto, S., Takamura, K., Sekine, T., … Takase, S. (2013). [Prospective electrocardiogram gated coronary 320-row area detector computed tomography angiography using low tube current scanning with full reconstruction]. Nihon Hoshasen Gijutsu Gakkai Zasshi, 69(3), 244–250. https://doi.org/10.6009/jjrt.2013_JSRT_69.3.244

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free