Abstract
Usability of medical devices is one of the main determining factors in preventing use errors in treatment and strongly correlates to patient safety and quality of treatment. This thesis demonstrates the usability testing and evaluation of a prototype for locomotor therapy of infants. Therefore, based on the normative requirements of the EN 62366, a concept combined of evaluation procedures and assessing methods was created to enable extensive testing and analysis of the different aspects of usability. On the basis of gathered information weak points were identified and appropriate measures were presented to increase the usability and operating safety of the locomotor prototype. The overall outcome showed an usability value of 77.4% and an evaluation score of 6.99, which can be interpreted as “satisfactory”.
Author supplied keywords
Cite
CITATION STYLE
Langthaler, S., Lassnig, A., Baumgartner, C., & Schröttner, J. (2016). Usability evaluation of a locomotor therapy device considering different strategies. In Current Directions in Biomedical Engineering (Vol. 2, pp. 67–69). Walter de Gruyter GmbH. https://doi.org/10.1515/cdbme-2016-0018
Register to see more suggestions