A feedback controller for ventilatory therapy

34Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A computerized system that uses feedback of end-tidal CO2 fraction (F ETCO2) to adjust minute volume of a ventilator has been developed and tested. The effectiveness and robustness of the controller were evaluated in five anesthetized dogs. The controlled responded to step-changes in the set-point for F ETCO2 by adjusting minute volume so that the F ETCO2 settled to the new set-point in less than 60 sec with less than 20% overshoot. The system exhibited suitable dynamic response to step-changes in set-point with loop gains as large as two times and as small as one-half the optimal value. The breath-to-breath variation in F ETCO2 values during prolonged periods of closed-loop controlled ventilation was smaller than the variation during periods of constant minute volume ventilation in three of five experiments. The controller generally maintained F ETCO2 within ±0.1 vol% of the set-point. A disturbance to the controlled system was produced by releasing an occlusion of a branch of the pulmonary artery. The controller always responded to this disturbance in a stable manner, returning the F ETCO2 to its desired value within 30 sec. Accurate control of arterial partial pressure of CO2(Pa CO2) will require modifications enabling the system to determine the relationship between F ETCO2 and Pa CO2. © 1985 Pergamon Press Ltd.

Cite

CITATION STYLE

APA

Chapman, F. W., Newell, J. C., & Roy, R. J. (1985). A feedback controller for ventilatory therapy. Annals of Biomedical Engineering, 13(5), 359–372. https://doi.org/10.1007/BF02407766

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free