Abstract
A computerized system that uses feedback of end-tidal CO2 fraction (F ETCO2) to adjust minute volume of a ventilator has been developed and tested. The effectiveness and robustness of the controller were evaluated in five anesthetized dogs. The controlled responded to step-changes in the set-point for F ETCO2 by adjusting minute volume so that the F ETCO2 settled to the new set-point in less than 60 sec with less than 20% overshoot. The system exhibited suitable dynamic response to step-changes in set-point with loop gains as large as two times and as small as one-half the optimal value. The breath-to-breath variation in F ETCO2 values during prolonged periods of closed-loop controlled ventilation was smaller than the variation during periods of constant minute volume ventilation in three of five experiments. The controller generally maintained F ETCO2 within ±0.1 vol% of the set-point. A disturbance to the controlled system was produced by releasing an occlusion of a branch of the pulmonary artery. The controller always responded to this disturbance in a stable manner, returning the F ETCO2 to its desired value within 30 sec. Accurate control of arterial partial pressure of CO2(Pa CO2) will require modifications enabling the system to determine the relationship between F ETCO2 and Pa CO2. © 1985 Pergamon Press Ltd.
Author supplied keywords
Cite
CITATION STYLE
Chapman, F. W., Newell, J. C., & Roy, R. J. (1985). A feedback controller for ventilatory therapy. Annals of Biomedical Engineering, 13(5), 359–372. https://doi.org/10.1007/BF02407766