Distinction and quantification of carry-over and sample interaction in gas segmented continuous flow analysis

40Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The formulae for calculation of carry-over and sample interaction are derived for the first time in this study. A scheme proposed by Thiers et al. (two samples of low concentration followed by a high concentration sample and low concentration sample) is verified and recommended for the determination of the carry-over coefficient. The derivation demonstrates that both widely used schemes of a high concentration sample followed by two low concentration samples, and a low concentration sample followed by two high concentration samples actually measure the sum of the carry-over coefficient and sample interaction coefficient. A scheme of three low concentration samples followed by a high concentration sample is proposed and verified for determination of the sample interaction coefficient. Experimental results indicate that carry-over is a strong function of cycle time and a weak function of ratio of sample time to wash time. Sample dispersion is found to be a function of sample time. Fitted equations can be used to predict the carry-over, absorbance and dispersion given sample times, and wash times for an analytical system. Results clearly show the important role of intersample air segmentation in reducing carry-over, sample interaction and dispersion.

Cite

CITATION STYLE

APA

Zhang, J. Z. (1997). Distinction and quantification of carry-over and sample interaction in gas segmented continuous flow analysis. Journal of Automatic Chemistry, 19(6), 205–212. https://doi.org/10.1155/S1463924697000254

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free