Abstract
Growing acceptance of the exponentially modified Gaussian (EMG) function as a good model for real chromatographic peaks has prompted a review of its use since 1983. This review covers the following areas: (1) equations derived from the EMG model, (2) studies of inherent errors in the quantitation of chromatographic peaks via use of the EMG model, (3) chromatographic applications since 1983, and (4) applications to flow injection analysis. The goals of this review are to provide a valuable resource to researchers already employing the EMG model, while also providing a foundation for prospective first time users. © 1991, Oxford University Press. All rights reserved.
Cite
CITATION STYLE
Jeansonne, M. S., & Foley, J. P. (1991). Review of the exponentially modified gaussian (EMG) function since 1983. Journal of Chromatographic Science, 29(6), 258–266. https://doi.org/10.1093/chromsci/29.6.258