Abstract
Reference intervals are essential for clinical laboratory test interpretation and patient care. Methods for estimating them are expensive, difficult to perform, often inaccurate, and nonreproducible. A computerized indirect Hoffmann method was studied for accuracy and reproducibility. The study used data collected retrospectively for 5 analytes without exclusions and filtering from a nationwide chain of clinical reference laboratories in the United States. The accuracy was assessed by the comparability of reference intervals as calculated by the new method with published peer-reviewed studies, and reproducibility was assessed by the comparability of 2 sets of reference intervals derived from 2 different data sets. There was no statistically significant difference between the calculated and published reference intervals or between the 2 sets of intervals that were derived from different data sets. A computerized Hoffmann method for indirect estimation of reference intervals using stored test results is proved to be accurate and reproducible. © American Society for Clinical Pathology.
Author supplied keywords
Cite
CITATION STYLE
Katayev, A., Balciza, C., & Seccombe, D. W. (2010). Establishing reference intervals for clinical laboratory test results: Is there a better way? American Journal of Clinical Pathology, 133(2), 180–186. https://doi.org/10.1309/AJCPN5BMTSF1CDYP
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.