Overcoming the effects of matrix interference in the measurement of urine protein analytes

22Citations
Citations of this article
76Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Using multiplex bead assays to measure urine proteins has a great potential for biomarker discovery, but substances in urine (the matrix) can interfere with assay measurements. By comparing the recovery of urine spiked with known quantities of several common analytes, this study demonstrated that the urine matrix variably interfered with the accurate measurement of low abundance proteins. Dilution of the urine permitted a more accurate measure of these proteins, equivalent to the standard dilution technique when the diluted analytes were above the limits of detection of the assay. Therefore, dilution can be used as an effective technique for overcoming urine matrix effects in urine immunoassays. These results may be applicable to other biological fluids in which matrix components interfere with assay performance. © the author(s), publisher and licensee Libertas Academica Ltd.

Cite

CITATION STYLE

APA

Taylor, T. P., Janech, M. G., Slate, E. H., Lewis, E. C., Arthur, J. M., & Oates, J. C. (2012). Overcoming the effects of matrix interference in the measurement of urine protein analytes. Biomarker Insights, 7, 1–8. https://doi.org/10.4137/bmi.s8703

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free