Spatial fuzzy c-means thresholding for semiautomated calculation of percentage lung ventilated volume from hyperpolarized gas and 1 H MRI

41Citations
Citations of this article
54Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: To develop an image-processing pipeline for semiautomated (SA) and reproducible analysis of hyperpolarized gas lung ventilation and proton anatomical magnetic resonance imaging (MRI) scan pairs. To compare results from the software for total lung volume (TLV), ventilated volume (VV), and percentage lung ventilated volume (%VV) calculation to the current manual “basic” method and a K-means segmentation method. Materials and Methods: Six patients were imaged with hyperpolarized 3 He and same-breath lung 1 H MRI at 1.5T and six other patients were scanned with hyperpolarized 129 Xe and separate-breath 1 H MRI. One expert observer and two users with experience in lung image segmentation carried out the image analysis. Spearman (R), Intraclass (ICC) correlations, Bland–Altman limits of agreement (LOA), and Dice Similarity Coefficients (DSC) between output lung volumes were calculated. Results: When comparing values of %VV, agreement between observers improved using the SA method (mean; R = 0.984, ICC = 0.980, LOA = 7.5%) when compared to the basic method (mean; R = 0.863, ICC = 0.873, LOA = 14.2%) nonsignificantly (p R = 0.25, p ICC = 0.25, and p LOA = 0.50 respectively). DSC of VV and TLV masks significantly improved (P < 0.01) using the SA method (mean; DSC VV = 0.973, DSC TLV = 0.980) when compared to the basic method (mean; DSC VV = 0.947, DSC TLV = 0.957). K-means systematically overestimated %VV when compared to both basic (mean overestimation = 5.0%) and SA methods (mean overestimation = 9.7%), and had poor agreement with the other methods (mean ICC; K-means vs. basic = 0.685, K-means vs. SA = 0.740). Conclusion: A semiautomated image processing software was developed that improves interobserver agreement and correlation of lung ventilation volume percentage when compared to the currently used basic method and provides more consistent segmentations than the K-means method. Level of Evidence: 3. Technical Efficacy: Stage 2. J. Magn. Reson. Imaging 2018;47:640–646.

Cite

CITATION STYLE

APA

Hughes, P. J. C., Horn, F. C., Collier, G. J., Biancardi, A., Marshall, H., & Wild, J. M. (2018). Spatial fuzzy c-means thresholding for semiautomated calculation of percentage lung ventilated volume from hyperpolarized gas and 1 H MRI. Journal of Magnetic Resonance Imaging, 47(3), 640–646. https://doi.org/10.1002/jmri.25804

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free