Determination of chlorine species by capillary electrophoresis – mass spectrometry

8Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The applicability of CZE with mass spectrometric detection for the determination of four chlorine species, namely chloride and three stable chlorine oxyanions, was studied. The main aspects of the proper selection of BGE and sheath liquid for the CE-MS determinations of anions with high mobility were demonstrated, pointing out the importance of pH and the mobility of the anion in the BGE. The possibility of using uncoated fused silica capillary and common electrolytes for the separation was shown and the advantage of using extra pressure at the inlet capillary end was also presented. The linear range was found to be 1–100 µg/mL for ClO3− and ClO4−, 5–500 µg/mL for ClO2−, and 25–500 µg/mL for Cl−, but the sensitivity can be greatly improved if larger sample volume is injected and electrostacking effect is utilized. The LOD for ClO3− in drinking water was 6 ng/mL, when very large sample volume was injected (10 000 mbar·s was applied).

Cite

CITATION STYLE

APA

Gaspar, A., Pesti, A., Szabo, M., & Kecskemeti, A. (2019). Determination of chlorine species by capillary electrophoresis – mass spectrometry. Electrophoresis, 40(20), 2637–2643. https://doi.org/10.1002/elps.201900138

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free