Abstract
An optical chemical sensor for the determination of nitrite based on incorporating methyltrioctylammonium chloride as an anionic exchanger on the triacetylcellulose polymer has been reported. The response of the sensor is based on the redox reaction between nitrite in aqueous solution and iodide adsorbed on sensing membrane using anion exchange phenomena. The sensing membrane reversibly responses to nitrite ion over the range of 6.52×10 -6 - 8.70×10 -5 mol L -1 with a detection limit of 6.05×10 -7 mol L -1 (0.03 μg mL -1) and response time of 6 min. The relative standard deviation for eight replicate measurements of 8.70×10 -6 and 4.34×10 -5 mol L -1 of nitrite was 4.4 and 2.5 %, respectively. The sensor was successfully applied for determination of nitrite in food, saliva and water samples.
Author supplied keywords
Cite
CITATION STYLE
Rastegarzadeh, S., & Kalantaripour, M. (2012). Development of an optical redox chemical sensor for nitrite determination. Quimica Nova, 35(4), 766–770. https://doi.org/10.1590/S0100-40422012000400021