Reliability of Water Quality Test Strips for Field Measurement of Nutrients Essential for Developing Mosquito Larvae

  • Isbell F
  • Mercer D
  • Iqbal M
N/ACitations
Citations of this article
17Readers
Mendeley users who have this article in their library.

Abstract

We tested the reliability of water quality test strips for field-analysis of nitrate and phosphate against laboratory-based standard methods for water samples drawn from a variety of natural and artificial sites used by mosquito larvae for development. Nitrate test strip measurements were compared with those from ion chromatographic analysis and an ion-selective electrode. Phosphate test strip measurements were compared to laboratory persulfate digestion followed by an ascorbic acid colorimetric assay. It was concluded that while the standard methods were the most reliable, nitrate-sensitive test strips were acceptable for measuring moderate nitrate concentrations, but phosphate-sensitive test strips were not reliable in measuring the forms of phosphorus available to mosquito larvae.

Cite

CITATION STYLE

APA

Isbell, F., Mercer, D., & Iqbal, M. (2006). Reliability of Water Quality Test Strips for Field Measurement of Nutrients Essential for Developing Mosquito Larvae. American Journal of Undergraduate Research, 5(2). https://doi.org/10.33697/ajur.2006.013

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free