Abstract
Aim. Comparison of the IgG-binding activity of recombinant Staphylococcal protein A with introduced C-terminal cysteine residue (SPA-Cys) or goat anti-human IgG antibodies (anti-IgG) after their immobilization on a gold sensor surface of surface plasmon resonance (SPR) spectrometer. Methods. SPA-Cys or anti- IgG were immobilized on a gold sensor surface to form two variants of a bioselective element of the immunosensor. SPR spectrometry was used for the detection of IgG-binding activity of the immobilized proteins. Results. The SPR sensor response to the immobilization of anti-IgG was more than two times higher than that to the immobilization of SPA-Cys. However, there is almost the double advantage for SPA-Cys in the number of immobilized molecules. Moreover, the bioselective element of the immunosensor based on SPA-Cys showed a much better capability of binding IgG than the bioselective element based on anti-IgG. Conclusions. Comparison of immobilization of SPA-Cys or anti-IgG on the sensor surface of SPR spectrometer, and the interactions of immobilized proteins with human IgG demonstrated obvious advantages of SPA-Cys.
Author supplied keywords
Cite
CITATION STYLE
Bakhmachuk, A. O., Gorbatiuk, O. B., Palyvoda, O. G., Dons’koi, B. V., Rachkov, A. E., & Soldatkin, A. P. (2016). Study on interactions of human IgG with immobilized anti-IgG or recombinant Staphylococcal protein A using surface plasmon resonance spectrometry. Biopolymers and Cell, 32(1), 54–60. https://doi.org/10.7124/bc.00090D
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.