Comparison of enzymatic recycling electrodes for measuring aminophenol: Development of a highly sensitive natriuretic peptide assay system

9Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

Several redox enzymes were examined for enzymatic/electrochemical-recycling systems in order to measure p-aminophenol (PAP) with high sensitivity. Glucose oxidase (GOD) and diaphorase (DI) worked well as catalysts for recycling electrode systems: these enzymes effectively reduced p-iminoquinone (PIQ), the electrochemically-oxidized form of PAP, and caused an enhancement in the electrochemical signals (anodic currents in the voltammogram and amperogram) by ∼100 fold. The lower detection limits for PAP were estimated to be 50 nM with the GOD system and 2 nM with the DI system. We combined the enzymatic-recycling electrode using DI with an enzyme immunoassay system to measure atrial natriuretic peptide (ANP), an important marker peptide hormone involved in heart diseases. ANPs from serum samples at ppt-levels were determined appropriately using the present assay system. 2008 © The Japan Society for Analytical Chemistry.

Cite

CITATION STYLE

APA

Mie, Y., Kowata, K., Hirano, Y., Niwa, O., & Mizutani, F. (2008). Comparison of enzymatic recycling electrodes for measuring aminophenol: Development of a highly sensitive natriuretic peptide assay system. Analytical Sciences, 24(5), 577–582. https://doi.org/10.2116/analsci.24.577

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free