Universal EDXRF method for multi-elemental determinations using fused bead specimens

5Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A systematic study on the EDXRF determination of multi-elements using fusion bead specimens and a universal calibration procedure was carried out. Beads of multi-elements were prepared by mixing MERCK multi-elemental standard solutions with the fusion flux. The elemental X-ray lines of K, Ca, Ba, Cr, Mn, Fe, Co, Ni, Cu and Zn were excited using a Ge secondary target and that of Cr, Mn, Fe, Co, Ni, Cu, Zn, Ga, Pb, Bi and Sr using a Rh primary target. The relative sensitivities of the elemental X-ray lines obtained using different excitation modes were calculated using a single bead. Using this relative sensitivity, the concentration of elements in other bead samples was determined. The average precision obtained was 3% (1 s) and the average deviation from expected values, calculated on the basis of sample preparation, was 6%. In order to further countercheck the validity of this methodology, a standard soil sample received from IAEA was analyzed.

Cite

CITATION STYLE

APA

Dhara, S., Sanyal, K., & Misra, N. L. (2020). Universal EDXRF method for multi-elemental determinations using fused bead specimens. Analytical Sciences, 36(1), 113–119. https://doi.org/10.2116/analsci.19P196