Towards a capacitive enzyme sensor for direct determination of organophosphorus pesticides: Fundamental studies and aspects of development

24Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

Abstract

The realisation of a miniaturised potentiometric enzyme biosensor is presented. The biosensor chip utilises the enzyme organophosphorus hydrolase (OPH) for the direct determination of pesticides. The transducer structure of the sensor chip consists of a pH-sensitive capacitive electrolyte-insulator- semiconductor (EIS) structure that reacts towards pH changes caused by the OPH-catalysed hydrolysis of the organophosphate compounds. The biosensor is operated versus a conventional Ag/AgCl reference electrode. Measurements were performed in the capacitance/voltage (C/V) and constant capacitance (ConCap) mode for the two different pesticides paraoxon and parathion. For the development of this new type of biosensor, different immobilisation strategies, influence of buffer composition and concentration, transducer material, detection limit, long-term stability and selectivity have been studied. © 2003 by MDPI.

Cite

CITATION STYLE

APA

Schöning, M. J., Arzdorf, M., Mulchandani, P., Chen, W., & Mulchandani, A. (2003). Towards a capacitive enzyme sensor for direct determination of organophosphorus pesticides: Fundamental studies and aspects of development. Sensors, 3(6), 119–127. https://doi.org/10.3390/s30600119