Objective: The aim of the present study was to evaluate a simplified method for FRC measurement. Design: Accuracy and precision of the method were assessed in a physical lung model; reproducibility was tested in 10 mechanically ventilated patients. In each patient FRC was measured at three PEEP levels. Setting: Post-operative intensive care unit in a university hospital. Measurements and results: Gas flow, CO2 concentration, and O2 concentration were measured during in- and expiration by pneumotachography, a mainstream capnometer and a sidertream O2-analyser. For FRC-measurement inspiratory O2 concentration was changed by 30%. FRC was determined as mean value of a N2 washout and N2 washin procedure. Evaluation of this method in a lung model shows a good correlation between FRC set in the lung model and FRC measured (FRC measured =1.028*FRC model+22.92 ml;r2=0.957;n=30). The mean difference was 4.4% of FRC-reference (range -8.4% to +21.7%). Duplicate determinations in 10 mechanically ventilated patients differed by an average of -2.7% (range -30.1% to +27.3%). Conclusion: Our results suggest that the proposed method can be used in daily clinical work. © 1993 Springer-Verlag.
CITATION STYLE
Fretschner, R., Deusch, H., Weitnauer, A., & Brunner, J. X. (1993). A simple method to estimate functional residual capacity in mechanically ventilated patients. Intensive Care Medicine, 19(7), 372–376. https://doi.org/10.1007/BF01724875