Rescue retrieval of AngioGuard embolic capture system after failure of capture sheath retrieval in carotid angioplasty and stenting

16Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

This report describes the cases of 2 patients with failed retrieval of an AngioGuard distal protection device after carotid stent placement with rescue retrieval by use of a vertebral catheter. An AngioGuard was deployed before carotid angioplasty and stent placement. Attempted removal of the device with the provided recapture sheath was unsuccessful. A 5F 120-cm vertebral catheter was used to recapture and remove the AngioGuard. This technique is a simple and readily available solution for the retrieval of failed removal of a protection device.

Cite

CITATION STYLE

APA

Daugherty, W. P., White, J. B., Cloft, H. J., & Kallmes, D. F. (2008). Rescue retrieval of AngioGuard embolic capture system after failure of capture sheath retrieval in carotid angioplasty and stenting. American Journal of Neuroradiology, 29(8), 1594–1595. https://doi.org/10.3174/ajnr.A1118