Abstract
We assayed rheumatoid factor by instrumental latex particle counting. The calibration curve ranged from 12.5 to 500 int. units/mL. Maximal within- and between-assay CVs were 5 and 11%, respectively. Analytical recoveries ranged from 92.4 to 108%, and the relation between results and dilutions was linear in the range of 15 to 400 int. units/mL. Correlation with an enzyme-linked immunoassay (ELISA, Cordia kit) was r = 0.934 (n = 58), with turbidimetry r = 0.825 (n = 100), and with the Waaler-Rose test r = 0.834 (n = 73). Of 260 blood donors, 95% gave a value < 10 kilo-int. units/L, which was taken as the upper normal limit. In a population of patients with rheumatoid arthritis (n = 47), 87.2% had rheumatoid factor > 10 kilo-int. units/L.
Cite
CITATION STYLE
Borque, L. A., Cambiaso, C. L., & Masson, P. L. (1987). Immunoassay of rheumatoid factor by latex particle counting. Clinical Chemistry, 33(5), 704–707. https://doi.org/10.1093/clinchem/33.5.704