Determination of trace metals in electrolytic copper by ICP OES and ICP-MS

19Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.

Abstract

The performance of axial view ICP OES and ICP-MS techniques were compared through the determination of As, Fe, Mn, Pb, Sb and Sn in electrolytic copper. Samples were prepared by two procedures: 1. Total dissolution with 5 % v/v HNO3 and 2. Dissolution with 1.4 % v/v HNO3 plus 2.0 % v/v H2SO4, followed by separation of the Cu by electrodeposition. The methods were applied for the analysis of standard copper solutions, using calibration against aqueous solutions with or without the addition of Cu. The results obtained by the two techniques were similar, but the separation of Cu from the sample matrix proved to be more adequate for practical reasons.