Commissioning and first results of the LNLS XAFS beamline

87Citations
Citations of this article
25Readers
Mendeley users who have this article in their library.
Get full text

Abstract

An X-ray absorption fine-structure spectroscopy beamline has been installed and commissioned at a bending-magnet source at LNLS. Three monochromators are available: a channel-cut, a double-crystal and a four-crystal set-up. They have been operated from 2500 up to 15000 eV, with a resolving power better than 5500 in the full range. Photon flux of the order of 108 photons s-1 up to 1010 photons s-1 has been attained. The experimental station is equipped with a table that can withstand a weight of 300 kg and track the vertical position of the beam with a 2.5 μm accuracy over a 120 mm stroke. The beamline has been fully characterized and the first results are presented.

Author supplied keywords

Cite

CITATION STYLE

APA

Tolentino, H., Cezar, J. C., Cruz, D. Z., Compagnon-Cailhol, V., Tamura, E., & Martins Alves, M. C. (1998). Commissioning and first results of the LNLS XAFS beamline. Journal of Synchrotron Radiation, 5(3), 521–523. https://doi.org/10.1107/S0909049597015264