Abstract
We present the application of a numerical model to estimate the mean glandular dose for ma- mmography. A General Electric mammography system, tube voltaje of 20-35 kV with increments of 1 kV, tube current of 10-250 mAs, anode/filter combinations molybdenum/molybdenum was used. Over a period of two months, data from 162 scans with craniocaudal and mediolateral oblique projections in both breasts were recorded. The thickness of the compressed breast ranged from 2.5 to 6.5 cm. The mean glandular dose for each craniocaudal projection ranged from 0.4 to 0.7 mGy for a thickness of 2.5 cm and 1.9 to 2.8 mGy for a thickness of 6.5 cm. All average values of mean glandular dose estimated by cranio-caudal projection were found to be below the international reference dose value of 3 mGy recommended by the American College of Radiology.
Cite
CITATION STYLE
Ramos N, O., & Villarreal U, M. (2009). DETERMINACIÓN DE UN MÉTODO DE CALCULO DE DOSIS GLANDULAR PROMEDIO EN EXAMENES DE MAMOGRAFIA CONVENCIONAL. Revista Chilena de Radiología, 15(4). https://doi.org/10.4067/s0717-93082009000500007
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.