Abstract

Protocol review is important to decrease the risk of patient injury and increase the consistency of CT image quality. A large volume of CT protocols makes manual review labor-intensive, error-prone, and costly. To address these challenges, we have developed a software system for automatically managing and monitoring CT protocols on a frequent basis. This article describes our experiences in the implementation and evaluation of this protocol monitoring system. In particular, we discuss various strategies for addressing each of the steps in our protocol-monitoring workflow, which are: maintaining an accurate set of master protocols, retrieving protocols from the scanners, comparing scanner protocols to master protocols, reviewing flagged differences between the scanner and master protocols, and updating the scanner and/or master protocols. In our initial evaluation focusing only on abdomen and pelvis protocols, we detected 309 modified protocols in a 24-week trial period. About one-quarter of these modified protocols were determined to contain inappropriate (i.e., erroneous) protocol parameter modifications that needed to be corrected on the scanner. The most frequently affected parameter was the series description, which was inappropriately modified 47 times. Two inappropriate modifications were made to the tube current, which is particularly important to flag as this parameter impacts both radiation dose and image quality. The CT protocol changes detected in this work provide strong motivation for the use of an automated CT protocol quality control system to ensure protocol accuracy and consistency.

References Powered by Scopus

FDA investigates the safety of brain perfusion CT

106Citations
N/AReaders
Get full text

Methods for CT automatic exposure control protocol translation between scanner platforms

31Citations
N/AReaders
Get full text

CT protocol review and optimization

30Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Exploring the Role of Artificial Intelligence in an Emergency and Trauma Radiology Department

39Citations
N/AReaders
Get full text

Tracking Patterns of Nonadherence to Prescribed CT Protocol Parameters

8Citations
N/AReaders
Get full text

AAPM Medical Physics Practice Guideline 1.b: CT protocol management and review practice guideline

8Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Grimes, J., Leng, S., Zhang, Y., Vrieze, T., & McCollough, C. (2016). Implementation and evaluation of a protocol management system for automated review of CT protocols. Journal of Applied Clinical Medical Physics, 17(5), 523–533. https://doi.org/10.1120/jacmp.v17i5.6164

Readers over time

‘16‘17‘18‘19‘20‘21‘22‘23‘24‘2502468

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 8

42%

Researcher 8

42%

Professor / Associate Prof. 2

11%

Lecturer / Post doc 1

5%

Readers' Discipline

Tooltip

Medicine and Dentistry 7

33%

Nursing and Health Professions 5

24%

Engineering 5

24%

Physics and Astronomy 4

19%

Save time finding and organizing research with Mendeley

Sign up for free
0