We evaluated two immunochromatographic assays (ICAs), Dainascreen HBsAg for detecting human hepatitis B surface antigen (HBsAg) and Dainascreen Ausab for detecting human hepatitis B surface antibody (anti-HBs) in human serum. The ICA systems are composed of a comb-shaped device that contains nitrocellulose strips on which complexes of HBsAg and anti-HBs can be visualized. The results can be read within 15 rain of incubation. The limit of detection for HBsAg was 3.1 ng/ml, and that for anti-HBs was 42 mlU/ml. Results of HBsAg detection agreed completely with those of conventional enzyme immunoassays (EIAs) and showed a 100% sensitivity (158 of 158 samples) and a 100% specificity (304 of 304 samples). The Dainascreen Ausab detected 184 of the 199 EIA-positive samples (sensitivity, 92.5%) and yielded 6 positive results among the 281 EIA-negative samples (specificity, 97.9%). The ICA systems are rapid and sensitive methods for detecting HBsAg and anti- HBs. They are low-cost systems that need no complex instrumentation for analysis and can be recommended for routine use in clinical microbiology laboratories.
Cite
CITATION STYLE
Sato, K., Ichiyama, S., Iinuma, Y., Nada, T., Shimokata, K., & Nakashima, N. (1996). Evaluation of immunochromatographic assay systems for rapid detection of hepatitis B surface antigen and antibody, Dainascreen HBsAg and Dainascreen Ausab. Journal of Clinical Microbiology, 34(6), 1420–1422. https://doi.org/10.1128/jcm.34.6.1420-1422.1996