In this paper, we present a stable silver-based surface plasmon resonance (SPR) sensor using a self-assembled monolayer (SAM) as a protection layer and investigated its efficiency in water and 0.01 M phosphate buffered saline (PBS). By simulation, silver-based SPR sensor has a better performance in field enhancement and penetration depth than that of a gold-based SPR sensor, which are 5 and 1.4 times, respectively. To overcome the instability of the bare silver film and investigate the efficiency of the protected layer, the SAM of 11-mercapto-1-undecanol (MUD) was used as a protection layer. Stability experiment results show that the protected silver film exhibited excellent stability either in pure water or 0.01 M PBS buffer. The sensitivity of the silver-based SPR sensor was calculated to be 127.26 deg/RIU (refractive index unit), measured with different concentrations of NaCl solutions. Further, a very high refractive resolution for the silver-based SPR sensor was found to be 2.207 × 10–7RIU, which reaches the theoretical limit in the wavelength of 632.8 nm for a SPR sensor reported in the literature. Using a mixed SAM of 16-mercaptohexadecanoic acid (MHDA) and a MUD layer with a ratio of 1:10, this immunosensor for the rabbit immunoglobulin G (IgG) molecule with a limit of detection as low as 22.516 ng/mL was achieved.
CITATION STYLE
Wang, G., Wang, C., Yang, R., Liu, W., & Sun, S. (2017). A sensitive and stable surface plasmon resonance sensor based on monolayer protected silver film. Sensors (Switzerland), 17(12). https://doi.org/10.3390/s17122777