How Fast is Your Detector? The Effect of Temporal Response on Image Quality

7Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

With increasing interest in high-speed imaging, there should be an increased interest in the response times of our scanning transmission electron microscope detectors. Previous works have highlighted and contrasted the performance of various detectors for quantitative compositional or structural studies, but here, we shift the focus to detector temporal response, and the effect this has on captured images. The rise and decay times of eight detectors’ single-electron response are reported, as well as measurements of their flatness, roundness, smoothness, and ellipticity. We develop and apply a methodology for incorporating the temporal detector response into simulations, showing that a loss of resolution is apparent in both the images and their Fourier transforms. We conclude that the solid-state detector outperforms the photomultiplier tube–based detectors in all areas bar a slightly less elliptical central hole and is likely the best detector to use for the majority of applications. However, using the tools introduced here, we encourage users to effectively evaluate which detector is most suitable for their experimental needs.

Cite

CITATION STYLE

APA

Mullarkey, T., Geever, M., Peters, J. J. P., Griffiths, I., Nellist, P. D., & Jones, L. (2023). How Fast is Your Detector? The Effect of Temporal Response on Image Quality. Microscopy and Microanalysis, 29(4), 1402–1408. https://doi.org/10.1093/micmic/ozad061

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free