Abstract
To measure the inhomogeneous 3D-strain fields present during inflation-extension testing of physiologically submerged micro-aneurysms, a Stereo Digital Image Correlation (StereoDIC) microscopy system is developed that revolves 15° stereo-angle cameras around a centrally-mounted target. Calibration is performed using submerged dot patterns and system accuracy verified using strain and deformation analyses for rigid body motions of speckle-patterned, micro-aneurysmal surrogates. In terms of the Green-Lagrange strain tensor and the 3D displacement fields, the results are stable even after 120 minutes, with maxima in both strain bias and strain standard deviation less than 2E-03 for all components, and micron-level displacement standard deviation.
Author supplied keywords
Cite
CITATION STYLE
Lane, B. A., Lessner, S. M., Vyavahare, N. R., Sutton, M. A., & Eberth, J. F. (2020). Null strain analysis of submerged aneurysm analogues using a novel 3D stereomicroscopy device. Computer Methods in Biomechanics and Biomedical Engineering, 23(8), 332–344. https://doi.org/10.1080/10255842.2020.1724974