Heart rate variability (HRV) measurements via ambulatory monitors have become common. We examined the validity of recording R-R intervals using the Polar V800™ compared to 12-lead electrocardiograms (ECG) among middle-aged (44.7±10.1years); overweight to obese (29.8±4.3 kg.m-2) adults (n = 25) with hypertension (132.3±12.2/ 84.3±10.2 mmHg). After resting for 5-min in the supine position, R-R intervals were simultaneously recorded using the Polar V800™ and the 12-lead ECG. Artifacts present in uncorrected (UN) R-R intervals were corrected with the Kubios HRV Premium (ver. 3.2.) automatic (AC) and threshold-based (TBC) correction, and manual correction (MC) methods. Intra-class correlation coefficients (ICC), Bland-Altman limits of agreement (LoA), and effect sizes (ES) were calculated. We detected 71 errors with the Polar V800™ for an error rate of 0.85%. The bias (LoAs), ES, and ICC between UN and ECG R-R intervals were 0.69ms (-215.80 to +214.42ms), 0.004, and 0.79, respectively. Correction of artifacts improved the agreeability between the Polar V800™ and ECG HRV measures. The biases (LoAs) between the AC, TBC, and MC and ECG R-R intervals were 3.79ms (-130.32 to +137.90ms), 1.16ms (-92.67 to +94.98ms), and 0.37ms (-41.20 to +41.94ms), respectively. The ESs of AC, TBC, and MC were 0.024, 0.008, and 0.002, and ICCs were 0.91, 0.95, and 1.00, respectively. R-R intervals measured using the Polar V800™ compared to 12-lead ECG were comparable in adults with hypertension, especially after the artifacts corrected by MC. However, TBC correction also yielded acceptable results.
CITATION STYLE
Cilhoroz, B., Giles, D., Zaleski, A., Taylor, B., Fernhall, B., & Pescatello, L. (2020). Validation of the Polar V800 heart rate monitor and comparison of artifact correction methods among adults with hypertension. PLoS ONE, 15(10). https://doi.org/10.1371/journal.pone.0240220
Mendeley helps you to discover research relevant for your work.