An open-source phase correction toolkit for transcranial focused ultrasound

  • Jin C
  • Moore D
  • Snell J
  • et al.
N/ACitations
Citations of this article
26Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The phase correction on transcranial focused ultrasound is essential to regulate unwanted focal point shift caused by skull bone aberration. The aim of the current study was to design and investigate the feasibility of a ray-based phase correction toolkit for transcranial focused ultrasound. The peak pressure at focal area was improved by 140.5 ± 7.0% on target I and 134.8 ± 19.1% on target II using proposed phase correction toolkit, respectively. A total computation time of 402.1 ± 24.5 milliseconds was achieved for each sonication. The designed ray-based phase correction software can be used as a lightweight toolkit to compensate aberrated phase within clinical environment.

Cite

CITATION STYLE

APA

Jin, C., Moore, D., Snell, J., & Paeng, D.-G. (2020). An open-source phase correction toolkit for transcranial focused ultrasound. BMC Biomedical Engineering, 2(1). https://doi.org/10.1186/s42490-020-00043-3

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free