Antithrombotic properties of hemofilter coated with polymer having a hydrophilic blood-contacting layer

4Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective: Extracorporeal circulation devices are coated with a biocompatible polymer coating agent (BPCA) that has a hydrophilic blood-contacting layer, but hemofilters are not. We aimed to investigate the antithrombotic properties of a BPCA-coated hemofilter. Methods: Four experiments using BPCA-coated circuits and non-coated hemofilters and four experiments using BPCA-coated circuits and BPCA-coated hemofilters were performed with whole human blood and compared by measuring the circuit pressure every 5 min, antithrombin activity every 40 min, and thrombin–antithrombin complex every 40 min, for a total of 240 min of recirculation. Results: The mean time required for the pressure at the inlet of the hemofilter to increase sharply was longer in BPCA-coated than in non-coated hemofilters (66 ± 11 min vs 25 ± 9 min, p < 0.01). The mean antithrombin activity value at 200 and 240 min of recirculation was significantly higher in the experiments with BPCA-coated versus non-coated hemofilters (43.3 ± 2.87 vs 33.3 ± 5.74, p = 0.04; 42.8 ± 3.59 vs 31.0 ± 5.35, p = 0.01, respectively); the antithrombin activity values at the other time points were not significantly different. Furthermore, all thrombin–antithrombin complex values in experiments with the BPCA-coated hemofilters achieved overrange at 80 min of recirculation, whereas those with the non-coated hemofilter achieved overrange at 40 min. Conclusion: This study suggests that BPCA-coated hemofilters can inhibit antithrombin consumption, contributing to antithrombotic effects in extracorporeal circulation circuits.

Cite

CITATION STYLE

APA

Tagaya, M., Hara, K., Takahashi, S., Nagoshi, S., Handa, H., Okano, S., & Murataka, T. (2019). Antithrombotic properties of hemofilter coated with polymer having a hydrophilic blood-contacting layer. International Journal of Artificial Organs, 42(2), 88–94. https://doi.org/10.1177/0391398818815480

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free