Coronary dissection due to overdilatation of an elastic membrane of the Multi-Link stent delivery system caused by balloon rupture

1Citations
Citations of this article
4Readers
Mendeley users who have this article in their library.

Abstract

A 65-year-old male with unstable angina underwent coronary angiography, which revealed a significant stenotic lesion in the right coronary artery. This narrowing was subsequently treated with the Multi-Link stent. During the balloon inflation associated with stent deployment, balloon rupture occurred and resulted in overdilatation of an elastic membrane in the stent delivery system. This, in turn, resulted in coronary dissection, which required treatment with further stenting.

Cite

CITATION STYLE

APA

Kuroda, N., Kobayashi, Y., De Gregorio, J., Kinoshita, T., Nameki, M., Yamamoto, Y., … Masuda, Y. (1999). Coronary dissection due to overdilatation of an elastic membrane of the Multi-Link stent delivery system caused by balloon rupture. Japanese Circulation Journal, 63(11), 912–913. https://doi.org/10.1253/jcj.63.912