An evaluation of osmolality measurement by freezing point depression using micro-amounts of sample

13Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

An evaluation of the Advanced micro-osmometer is presented. This instrument has been shown to have an excellent analytical precision (within-run CV = 0.59%, between-day CV = 0.58%). It is accurate over an analytical range of 0-2000 mmol/kg of osmolality shown by linearity studies and split sample correlations against vapour pressure osmometry, freezing point osmometry and an external quality assurance programme. Analytical errors due to operator technique are almost eliminated because of good instrument design. Preliminary results on whole-blood osmolality are included. The required sample size of 20 μl permits osmolality measurements on most clinical samples. It is concluded that the Advanced micro-osmometer satisfies laboratory requirements. © 1989, Taylor & Francis Ltd.

Cite

CITATION STYLE

APA

Koumantakis, G., & Wyndham, L. E. (1989). An evaluation of osmolality measurement by freezing point depression using micro-amounts of sample. Journal of Automatic Chemistry, 11(2), 80–83. https://doi.org/10.1155/S1463924689000167