Abstract
This study presents a passive, self-driven microfluidic device which can rapidly identify influenza A (HlNl) virus by using reverse transcription loop-mediated isothermal amplification (RT-LAMP). H1N1-specific aptamer conjugated magnetic beads were used to capture virus and the RT-LAMP product could be detected by visual coloration. The entire diagnosis could be automatically performed by capillary forces through a new polydimethylsiloxane (PDMS) surface treatment in cooperation with hydrophobic micro valves. The entire assay and its feasibility was verified. This is the first time that a simple, self-driven passive microfluidic chip was demonstrated which could perform the entire process including sample pretreatment, RT-LAMP application and influenza A diagnosis.
Author supplied keywords
Cite
CITATION STYLE
Ma, Y. D., Chang, W. H., Wang, C. H., Ma, H. P., Huang, P. C., & Lee, G. B. (2017). An integrated passive microfluidic device for rapid detection of influenza a (H1N1) virus by reverse transcription loop-mediated isothermal amplification (RT-LAMP). In TRANSDUCERS 2017 - 19th International Conference on Solid-State Sensors, Actuators and Microsystems (pp. 722–725). Institute of Electrical and Electronics Engineers Inc. https://doi.org/10.1109/TRANSDUCERS.2017.7994150