Clinical field-testing of ATLAS prosthetic system for trans-femoral amputees

14Citations
Citations of this article
56Readers
Mendeley users who have this article in their library.

Abstract

The ATLAS prosthetic system was tested on 47 trans-femoral amputees in two tropical areas. Although the craftsmanship was not acceptable, 46% (16/35) were malaligned many with the foot dorsiflexed, the user compliance was good (91%, 32/35), and the satisfaction rate high (71%, 25/35). However, after 7-10 months failures occurred in 83% (29/35) mostly because of problems with the ATLAS knee mechanism. A modification was tested and 80% (12/15) had again failed at the knee after 6 (2-12) months. The ATLAS system in its current design should be abandoned for use for trans-femoral amputees.

Cite

CITATION STYLE

APA

Jensen, J. S., & Raab, W. (2003). Clinical field-testing of ATLAS prosthetic system for trans-femoral amputees. Prosthetics and Orthotics International, 27(1), 55–62. https://doi.org/10.3109/03093640309167977