Abstract

The purpose of this paper is to report a salvage maneuver for accidental coverage of both renal arteries during endovascular aneurysm repair (EVAR) of an infrarenal abdominal aortic aneurysm (AAA). A 72-year-old female with a 6 cm infrarenal abdominal aortic aneurysm was treated by endovascular means with a standard bifurcated graft. Upon completing an angiogram, both renal arteries were found to be accidentally occluded. Through a left percutaneous brachial approach, the right renal artery was catheterized and a chimney stent was deployed; however this was not possible for the left renal artery. A retroperitoneal surgical approach was therefore carried out with a retrograde chimney stent implanted to restore blood flow. After three months, both renal arteries were patent and renal function was not different from the baseline. Both endovascular with percutaneous access via the brachial artery and open retroperitoneal approaches with retrograde catheterization are feasible rescue techniques to recanalize the accidentally occluded renal arteries during EVAR.

Cite

CITATION STYLE

APA

Bracale, U. M., Giribono, A. M., Vitale, G., Narese, D., Santini, G., & del Guercio, L. (2014). Accidental Coverage of Both Renal Arteries during Infrarenal Aortic Stent-Graft Implantation: Cause and Treatment. Case Reports in Vascular Medicine, 2014, 1–4. https://doi.org/10.1155/2014/710742

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free