System Accuracy and Interference Evaluation of a New Glucose Dehydrogenase-Based Blood Glucose Meter for Patient Self-Testing

1Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

New European medical device regulations require the performance of postmarketing surveillance evaluations for blood glucose meters (BGMs). We conducted an ISO15197:2015-conform system performance evaluation with the approved glucose dehydrogenase (GDH)-based Wellion NEWTON BGM. One hundred subjects were enrolled into the study (44 female, 56 male, 43 healthy subjects, 23 type 1 diabetes, 34 type 2 diabetes, age: 53.7 ± 15.8 years). In addition, manipulated heparinized whole blood was used for a laboratory interference test with ten selected substances (interference definition: substance-induced bias > 10%). The mean absolute relative difference (MARD) was 4.7%, and 100% of the values were in zones A (99.7%) and B (0.3%), respectively, of the consensus error grid. Interference was observed with xylose only, which is a known interfering substance for GDH-based BGMs.

Cite

CITATION STYLE

APA

Pfützner, A., Kalasauske, D., Hanna, M., Sachsenheimer, D., Raab, G., Weissenbacher, S., & Thomé, N. (2025). System Accuracy and Interference Evaluation of a New Glucose Dehydrogenase-Based Blood Glucose Meter for Patient Self-Testing. Journal of Diabetes Science and Technology, 19(2), 431–435. https://doi.org/10.1177/19322968231201862