Abstract
New European medical device regulations require the performance of postmarketing surveillance evaluations for blood glucose meters (BGMs). We conducted an ISO15197:2015-conform system performance evaluation with the approved glucose dehydrogenase (GDH)-based Wellion NEWTON BGM. One hundred subjects were enrolled into the study (44 female, 56 male, 43 healthy subjects, 23 type 1 diabetes, 34 type 2 diabetes, age: 53.7 ± 15.8 years). In addition, manipulated heparinized whole blood was used for a laboratory interference test with ten selected substances (interference definition: substance-induced bias > 10%). The mean absolute relative difference (MARD) was 4.7%, and 100% of the values were in zones A (99.7%) and B (0.3%), respectively, of the consensus error grid. Interference was observed with xylose only, which is a known interfering substance for GDH-based BGMs.
Author supplied keywords
Cite
CITATION STYLE
Pfützner, A., Kalasauske, D., Hanna, M., Sachsenheimer, D., Raab, G., Weissenbacher, S., & Thomé, N. (2025). System Accuracy and Interference Evaluation of a New Glucose Dehydrogenase-Based Blood Glucose Meter for Patient Self-Testing. Journal of Diabetes Science and Technology, 19(2), 431–435. https://doi.org/10.1177/19322968231201862