Fast kinetic measurements and on‐line dilution by flow injection cytometry

14Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

An improved flow injection cytometry (FIC) system suitable for fast cellular kinetic measurements and on‐line dilution is described. The instrument allows for measurements from 1.2 s (± 0.05 s) after the initiation of mixing, up to any time thereafter. Crucial factors in determining fast kinetic measurements, such as the displacement volume for sample introduction, rate of transport to cytometric interrogation point, and the mixing speed are evaluated and discussed. By varying the volume aspirated, this instrument can facilitate the dilution of cells and/or reagent over a range of one order of magnitude. The fast kinetic and on‐line dilution capabilities were demonstrated by on‐line staining of DNA in trout erythrocytes with 4′, 6‐diamidino‐2‐phenylindole (DAPI). The utility of the instrument for measurement of enzyme kinetics was illustrated using human lymphocytes, measuring the glutathione S‐transferase (GST) catalyzed reaction between monochlorobimane (MCB) and glutathione (GSH). © 1994 Wiley‐Liss, Inc. Copyright © 1994 Wiley‐Liss, Inc.

Cite

CITATION STYLE

APA

Lindberg, W., Scampavia, L. D., Ruzicka, J., & Christian, G. D. (1994). Fast kinetic measurements and on‐line dilution by flow injection cytometry. Cytometry, 16(4), 324–330. https://doi.org/10.1002/cyto.990160406