Experiment and theory combine to produce a practical negative ion calibration set for collision cross-section determinations by travelling-wave ion-mobility mass spectrometry

24Citations
Citations of this article
41Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

RATIONALE There are relatively few cross-section measurements for negatively charged ions. Available calibrants provide sufficient cross-section coverage for the 390 Å2 to 641 Å2 and 1174 Å2 to 3395 Å2 ranges. This is not particularly well suited for determining the collision cross-sections of smaller ions, such as small peptides. METHODS Molecular mechanics/molecular dynamics (MM/MD) simulations, coupled with simulated annealing, were used to find the low-energy molecular conformations of polystyrene (PS) oligomers of length 3-9 (singly deprotonated) and 5-13 (doubly deprotonated). The trajectory method in MOBCAL was employed to derive their respective collision cross-sections, Ω. A calibration plot relating corrected Ω values to drift times in a Waters Synapt G2 mass spectrometer was used to predict the Ω values for the -2 to -6 charge states of dT10 DNA. RESULTS The in silico design of a reliable negative ion calibration set for ion mobility spectrometry successfully resulted in the use of α,ω-carboxy-terminated PS oligomers to determine the collision cross-sections of negatively charged ions in the range 132-388 Å2. All charge states of dT10 DNA were predicted to within 3% of the referenced values for these ions. CONCLUSIONS α,ω-Carboxy-terminated PS oligomers were found to be an excellent choice to calibrate ion mobility spectrometers to obtain cross-sections for moderately sized ions. Oligomers with fewer, or weaker, interactions among the internal side chains (like poly(ethylene glycol) oligomers) tend to have a wide range of low-energy molecular conformations resulting in large standard deviations in their theoretically predicted collision cross-sections. © 2012 John Wiley & Sons, Ltd.

Cite

CITATION STYLE

APA

Hamilton, J., Renaud, J., & Mayer, P. (2012). Experiment and theory combine to produce a practical negative ion calibration set for collision cross-section determinations by travelling-wave ion-mobility mass spectrometry. Rapid Communications in Mass Spectrometry, 26(14), 1591–1595. https://doi.org/10.1002/rcm.6266

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free