Abstract
Introduction: Optical three-dimensional scanning devices can produce geometrically accurate, high-resolution models of patients suitable for clinical use. This article describes the use of a metrology-grade structured light scanner for the design and production of radiotherapy medical devices and synthetic water-equivalent computer tomography images. Methods: Following commissioning of the device by scanning objects of known properties, 173 scans were performed on 26 volunteers, with observations of subjects and operators collected. Results: The fit of devices produced using these scans was assessed, and a workflow for the design of complex devices using a treatment planning system was identified. Conclusions: Recommendations are provided on the use of the device within a radiation oncology department.