A fast fluorimetric flow injection method to determine ibuprofen

3Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

Abstract

The native fluorescence of ibuprofen was enhanced when a host-guest complex of the analyte with β cyclodextrin (β-CD) was formed. So, based on this fact, a flow injection method to determine ibuprofen in different pharmaceutical preparations with fluorescence detection (λem 287 nm, λexc 223 nm) was developed. The calibration curve was linear over the range 6.00 -60.0 mg L-1 of ibuprofen and the detection limit (LOD) for S/N=3 was 4.5 mg L-1. The relative standard deviation was 1.2% and the sample throughput 240 h-1. The method was validated by comparing the proposed and the official method to commercial preparation samples. © 2006 Sociedade Brasileira de Química.

Cite

CITATION STYLE

APA

Lista, A. G., Palomeque, M. E., & Fernández Band, B. S. (2006). A fast fluorimetric flow injection method to determine ibuprofen. Journal of the Brazilian Chemical Society, 17(7), 1428–1431. https://doi.org/10.1590/S0103-50532006000700033