Measurement of neuromuscular blockade - A comparison between a new 'homemade' force displacement transducer and the accelerometer

5Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Train-of-four (TOF) ratios are compared between a new homemade force pressure transducer and a commercially available accelerometer. Twenty adult patients were studied and 600 simultaneous TOF ratio pairs were recorded during anaesthesia with varying degrees of muscle relaxation. Laboratory testing revealed that the output from the homemade device was linear up to 900 g and that there was a linear relationship between the output from the new device and the accelerometer (y=0.802x + 5.167, R=0.943, t=69, P<0.0001). Although the homemade device underestimates the accelerometer TOF ration by 20%, it is a reliable and economic alternative to commercially available products for measuring TOF ratios.

Cite

CITATION STYLE

APA

Newell, S., & Brimacombe, J. (1995). Measurement of neuromuscular blockade - A comparison between a new “homemade” force displacement transducer and the accelerometer. Anaesthesia and Intensive Care, 23(2), 203–205. https://doi.org/10.1177/0310057x9502300215