Abstract
New generation oxygenators incorporate arterial line filtration either sequential to, or directly in, the gas exchange module. This unique design may affect gas exchange performance by altering the operational characteristics of the device. The present study was designed to evaluate three oxygenators in a clinical setting using a goal-directed perfusion algorithm during cardiopulmonary bypass (CPB). After institutional review board approval, 60 adult patients undergoing cardiac surgery for acquired heart disease were matched for disease state and body size into three groups based on oxygenator type: Terumo SX18⢠(n = 20), Terumo FX15⢠(n = 20), and LivaNova Inspire6F⢠6 Dual (n = 20). An external arterial line filter was used with the FX15, whereas the SX18 and Inspire6F had integrated arterial filters. All perfusion, anesthetic and postoperative care management was standardized using institutional goal-directed patient management processes. Data were collected and stored according to quality improvement guidelines. There were no differences in demographics or type of surgical procedure performed among groups. The Inspire6F patients required lower fraction of inspired oxygen values as compared to the SX18 (67.9% ± 6.2% vs. 75.4% ± 6.5%, p
Author supplied keywords
Cite
CITATION STYLE
Stammers, A. H., Miller, R., Francis, S. G., Fuzesi, L., Nostro, A., & Tesdahl, E. (2017). Goal-Directed Perfusion Methodology for Determining Oxygenator Performance during Clinical Cardiopulmonary Bypass. Journal of Extra-Corporeal Technology, 49(2), 81–92. https://doi.org/10.1051/ject/201749081