Sorption-spectrophotometric method for the determination of Pd(II) in aqueous solutions

7Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

The reaction of Pd(II) with 1,8-dihydroxy-2-(pyrazol-5-ylazo)-naphthalene- 3-6-disulphonic acid (PACA) sorbed onto Dowex 1-X8 ion-exchange resin was investigated with the aim of developing an absorption-spectrophotometric analytical method for the determination of low Pd(II) concentrations in water. The immobilized reagent formed a 1:1 complex with Pd(II) having an absorption maximum at 650 nm. Parameters, such as pH, wavelength and contact time were optimised for a given amount of the sorbed reagent. The linearity range of absorbance vs. Pd(II) concentration extended from 5×10-6 - 5×10-5 M, 5×10-7 - 5×10-6 and 2.5×10-8 - 2.5×10-7 M when using 10, 100 and 200 ml of sample solution, respectively. With a 200 ml sample, the detection limit was 2.5×10-7 M Pd(II). Most metals, except Cu(II), did not interfere when present in up to 100 times the concentration of Pd(II).

Cite

CITATION STYLE

APA

Vasić, V., Savić, J., & Vukelić, N. (2004). Sorption-spectrophotometric method for the determination of Pd(II) in aqueous solutions. Journal of the Serbian Chemical Society, 69(4), 309–317. https://doi.org/10.2298/JSC0404309V