Evaluation of nitrogen dioxide scavengers during delivery of inhaled nitric oxide

4Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We have analysed the ability of three nitrogen dioxide absorbing materials (soda lime, noXon and zeolite) to act as nitrogen dioxide scavengers during delivery of inhaled nitric oxide. Different mixtures of gas were produced in a ventilator (Servo Ventilator 300) and passed through an inspiratory tube. Concentrations of nitrogen dioxide and nitric oxide were measured in the distal part of the tube, with and without the gas having passed through a canister containing the different filter materials. Our findings indicated that nitrogen dioxide was absorbed effectively by all filter materials but that there was re-formation of nitrogen dioxide from nitric oxide and oxygen in or immediately after the canister. This initial production of nitrogen dioxide was very rapid and could not be prevented by the use of scavengers. Thus soda lime and zeolite had no practical effect as scavengers in this delivery system, and the effect of noXon was very slight.

Cite

CITATION STYLE

APA

Lindberg, L., & Rydgren, G. (1998). Evaluation of nitrogen dioxide scavengers during delivery of inhaled nitric oxide. British Journal of Anaesthesia, 81(3), 404–408. https://doi.org/10.1093/bja/81.3.404