Abstract
Background: The new TomoDirect™ modality offers a non-rotational option with discrete beam angles. We have investigated this mode for TBI with the intention to test the feasibility and to establish it as a clinical routine method. Special foci were directed onto treatment planning, dosimetric accuracy and practical aspects. Patients and methods: TBI plans were calculated with TomoDirect™ for a Rando™ phantom and all patients with an intended fractionated total body irradiation between November 2013 and May 2014 (n=8). Finally, four of these patients were irradiated with TomoDirect™. Additionally we studied variations in the modulation factor, pitch, field width of Y-jaws and dose grid during optimization. Dose measurements were performed using thermoluminescent rods in the Rando™ phantom, with the Delta4® and with ionization chambers in a solid water phantom. Results: For all eight calculated plans with a prescribed dose of 12Gy Dmean was 12.09-12.33Gy (12,25±0.08Gy), D98 11.2-11.6Gy (11.45±0.12Gy) and D2 12.6-13.1Gy (12.94±0.13Gy). Dmean of inner lungs was 8.73±0.22Gy on the left side and 8.69±0.27Gy on the right side. Conclusion: TBI with TomoDirect™ allows a superior homogeneity compared to conventional methods, where lung blocks are widely accepted. The treatment is performed only in supine position and is robust and comfortable for the patient.
Author supplied keywords
Cite
CITATION STYLE
Salz, H., Bohrisch, B., Howitz, S., Banz, N., Weibert, K., Wiezorek, T., & Wendt, T. G. (2015). Intensity-modulated Total Body Irradiation (TBI) with TomoDirectTM. Radiation Oncology, 10(1). https://doi.org/10.1186/s13014-015-0362-3
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.