Lung Function Monitoring; A Randomized Agreement Study

  • Berntsen S
  • Stølevik S
  • Mowinckel P
  • et al.
4Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

Abstract

OBJECTIVE: To determine the agreement between devices and repeatability within devices of the forced expiratory volume in 1 s (FEV1), forced vital capacity (FVC), peak expiratory flow (PEF) and forced expiratory flow at 50% of FVC (FEF50) values measured using the four spirometers included in the study. METHODS: 50 (24 women) participants (20-64 years of age) completed maximum forced expiratory flow manoeuvres and measurements were performed using the following devices: MasterScreen, SensorMedics, Oxycon Pro and SpiroUSB. The order of the instruments tested was randomized and blinded for both the participants and the technicians. Re-testing was conducted on a following day within 72 hours at the same time of the day. RESULTS: The devices which obtained the most comparable values for all lung function variables were SensorMedics and Oxycon Pro, and MasterScreen and SpiroUSB. For FEV1, mean difference was 0.04 L (95% confidence interval; -0.05, 0.14) and 0.00 L (-0.06, 0.06), respectively. When using the criterion of FVC and FEV1

Cite

CITATION STYLE

APA

Berntsen, S., Stølevik, S. B., Mowinckel, P., Nystad, W., & Stensrud, T. (2016). Lung Function Monitoring; A Randomized Agreement Study. The Open Respiratory Medicine Journal, 10(1), 51–57. https://doi.org/10.2174/1874306401610010051