Abstract
OBJECTIVE: To determine the agreement between devices and repeatability within devices of the forced expiratory volume in 1 s (FEV1), forced vital capacity (FVC), peak expiratory flow (PEF) and forced expiratory flow at 50% of FVC (FEF50) values measured using the four spirometers included in the study. METHODS: 50 (24 women) participants (20-64 years of age) completed maximum forced expiratory flow manoeuvres and measurements were performed using the following devices: MasterScreen, SensorMedics, Oxycon Pro and SpiroUSB. The order of the instruments tested was randomized and blinded for both the participants and the technicians. Re-testing was conducted on a following day within 72 hours at the same time of the day. RESULTS: The devices which obtained the most comparable values for all lung function variables were SensorMedics and Oxycon Pro, and MasterScreen and SpiroUSB. For FEV1, mean difference was 0.04 L (95% confidence interval; -0.05, 0.14) and 0.00 L (-0.06, 0.06), respectively. When using the criterion of FVC and FEV1